HEIDI grids on an optical bench, illuminated with a Helium Neon laser. The
reflected and transmitted beams are complementary to each other. The black
bars, seen orthogonal to the slits, are interference fringes
resulting from the small, deliberate misalignments between the grids. The
anticipated diffraction effects were, thus, visually confirmed.
This test was the precursor to the complete laser calibration of the HEIDI
telescope described in the abstract below.
Quantitative Characterization of the X-Ray Imaging
Capability of Rotating Modulation Collimators
Using Laser Light
- C. J. Gaither, III*
- NAS/NRC Fellow
- Goddard Space Flight Center, Code 682
- E. J. Schmahl
- University of Maryland
- C. J. Crannell, B. R. Dennis, F. L. Lang**, L. E. Orwig
- C. N. Hartman
- Goddard Space Flight Center, Code 735
- G. J. Hurford
- California Institute of Technology
Accepted by Applied Optics, August, 1996.
Abstract
We have developed a method for making quantitative characterizations of bi-grid
Rotating Modulation Collimators (RMCs) that are used in a Fourier
Transform X-ray imager. With appropriate choices of the collimator
spacings, this technique can be implemented with a beam-expanded HeNe
laser to simulate the plane wave produced by a point source at
infinity even though the RMCs are diffraction-limited at the HeNe
wavelength of 632.8 nm. The expanded beam passes through the grid
pairs at a small angle with respect to their axis of rotation, and the
modulated transmission through the grids as the RMCs rotate is
detected using a photomultiplier tube. In addition to providing a
quantitative characterization of the RMCs, the method also produces a
point response function and provides an end-to-end check of the
imaging system. We applied our method to the RMCs on the High Energy
Imaging Device (HEIDI) balloon payload in its preflight configuration.
We have computed the harmonic ratios of the modulation time profile
from the laser measurements and compared them with theoretical
calculations including the diffraction effects on irregular grids.
Our results indicate the 25'' X-ray imaging optics on HEIDI
are capable of achieving images near the theoretical limit and are not
seriously compromised by imperfections in the grids.
-
Key Words:
- Fourier transform imaging, imaging systems, laser
applications, optical testing, rotating modulation collimators, X-ray imagers.
* Currently at Analytic Services, Inc., Arlington, VA.
**Also Department of Physics, The Catholic University of America, Washington, DC.
Postscript version of published paper.
Figs 3a and 3b
Figs 7c and 7d
Fig 11
Last revised November 26, 1996.